Noise Injection into Cooling Pipe

 NEW RESULTS (25/04/02)

 
Module Position
K5-302 O10  low
K5-402 O11  high
K5-400 O12  low 
K5-303 O13  high

Ned Spencer's grounding scheme

Noise Injection over a transformer into the cooling pipe. In between the transformer and the cooling pipe is a resistor, to measure the voltage drop, from which the current conduction could be calculated.

Thermistor Temperature: 40 deg. C

Macros to analyse the data are provided by Joern.

1.) Measurements with sinus-waveformgenerator Wavetek 395 (as a loan from the barrel systemtest)

a.)     different Frequencies  4Mhz, 10Mhz, 20Mhz, 30Mhz and 40Mhz
         the current through the cooling pipe was for all measurements set to 50mA peak to peak.

        Difference in inputnoise

The highest increment for the inputnoise could be found for 10Mhz, but also for 20MHz and 30Mhz an increment could be observed. For 4MHz and 40Mhz the increment is small.
b.)    different currents:  12mA, 24mA, 36mA, 48mA and 60mA

        @ 10MHz                     @30MHz

        !!!! 1Vpp is equivalent to 6mA !!!
 
 
 
 

2.) Measurements with HP-Pulser 8012B

Because the Wavetek 395 was needed at the barrel systemtest and we could not get a working wave-generator from the electronic pool, a hp-pulser was used to inject the current into the pipe. The time dependece of the current through the pipe has a small similarity with a sinusfunction ........

 

However, the current was not changed for the following measurements

Difference without noiseinjection / with noiseinjection

Conclusions