Indium doped samples

 

Two detectors were fabricated with 500$\mu$m thick SI GaAs doped with indium [5]. These were tested before and after an exposure of 1.4 $\times$ 1014 24GeV/c protons/cm2. The reverse current of the detectors before irradiation was 30nA/mm2, less than for the 500$\mu$m thick SI-U GaAs detectors. After the irradiation the value of Vfd was reduced to only 220V, compared to 340V for the undoped sample. The current at the applied bias of Vfd was 0.4$\mu$A (57nA/mm2) which was higher than that of the undoped sample which was 0.22$\mu$A. The charge collection efficiency for the indium doped samples was less that that of the undoped samples before irradiation by 20%. After the irradiation the signal fell to 10% (6650 electrons) compared to 17% (11305 electrons), respectively, for an applied bias of 1000V.